Looking at the RF distribution system for corner and end stations there are only 3 units which show an error:
-
The RF divider for the 40MHz indicates a problem with the power supply. After swapping some cables this was traced to the RF Amp concentrator where channel 1 seems busted.
-
The RF amplifier for TCS where the field cable is missing.
-
The 90MHz LO for ASAIR_B which indicates no signal.
There is also an error associated with the green WFS demodulators where the hardware for the readbacks is still in the works.
The 90MHz demod problem was traced to a crosswiring of the demod readbacks of ASAIR_A and ASAIR_B. Decided to swap the DB9 cables.