Reports until 12:44, Friday 10 April 2015
LHO VE
kyle.ryan@LIGO.ORG - posted 12:44, Friday 10 April 2015 - last comment - 14:00, Friday 10 April 2015(17811)
~1230 hrs. local -> Valved-in HAM6 ion pump
HAM6 being pumped by both turbo and ion pump for now
Comments related to this report
kyle.ryan@LIGO.ORG - 14:00, Friday 10 April 2015 (17815)
Ion pump controller "protect" feature shut down after 15 minutes -> Can't handle relative high pressure (3000V @ 0.05A 150W unit) -> Valved-out ion pump from HAM6 -> Restarted ion pump -> OK now