Displaying report 1-1 of 1.
Reports until 16:07, Monday 13 April 2015
LHO VE
kyle.ryan@LIGO.ORG - posted 16:07, Monday 13 April 2015 (17851)
HAM6 almost recovered
Kyle, Gerardo 

We switched HAM5/6 annulus pumping over to the small annulus ion pumps and removed the pump carts -> The 500 l/s "main" ion pump was valved into HAM6 but we are leaving the turbo pump pumping in parallel with it over night 


- Commissioners -

HAM6 high voltages components (picomotors, PZT and fast shutter) can be left energized/enabled without monitoring HAM6 pump curve now.
Displaying report 1-1 of 1.