Since DC lock was achieved at 11:09 (after starting to try locking at 9:00 this morning), I've set the intent bit to Undisturbed. Injections are ongoing, the range isn't great and there have been earlier lock losses, but for now I think we can call an official start to the mini-run at LHO.
Andy, all: yep we had one calibration sweep in place during the first lock, apologies. From now on, we'll only allow for hardware injections when the intent bit is set.
Let us know if there is an interval you require (e.g. 2048s?) betwen injections.
We lost lock ~20 minutes ago. I've reset the intent bit. Looks like a rung up violin mode on ITMY. Jeff is working on damping it.