Displaying report 1-1 of 1.
Reports until 12:30, Monday 04 May 2015
H1 PSL
patrick.thomas@LIGO.ORG - posted 12:30, Monday 04 May 2015 (18208)
Ops PSL check
Laser Status: 
SysStat is good
Front End power is 31.4W (should be around 30 W)
FRONTEND WATCH is RED
HPO WATCH is RED

PMC:
It has been locked 0 day, 23 hr 54 minutes (should be days/weeks)
Reflected power is 2.0 Watts and PowerSum = 24.3 Watts.
(Reflected Power should be <= 10% of PowerSum)

FSS:
It has been locked for 5 h and 4 min (should be days/weeks)
TPD[V] = ~1.1V (should be 1.3 - 1.6)

ISS:
The diffracted power is around 8.5% (should be around 7%)
Last saturation event was 22 h and 57 minutes ago (should be days/weeks)
Displaying report 1-1 of 1.