Displaying report 1-1 of 1.
Reports until 12:33, Thursday 07 May 2015
H1 General (CDS)
travis.sadecki@LIGO.ORG - posted 12:33, Thursday 07 May 2015 - last comment - 20:38, Thursday 07 May 2015(18302)
Updated LSC PD Null script

At Kiwamu's request, I have updated his PD Null script, pdOffsetNull_ver2.py, located in opt/rtcds/userapps/release/lsc/h1/scripts, to include the balance of the LSC PDs.  The list of PDs zeroed by this script is now:

'LSC-POPAIR_B_RF18',
'LSC-POP_A_RF9',
'LSC-POP_A_RF45',
'LSC-POPAIR_A_RF9',
'LSC-POPAIR_A_RF45',
'LSC-POPAIR_B_RF90',
'LSC-ASAIR_B_RF18',
'LSC-ASAIR_B_RF90',
'LSC-REFLAIR_A_RF9',
'LSC-REFLAIR_A_RF45',
'LSC-REFLAIR_B_RF27',
'LSC-REFLAIR_B_RF135',
'LSC-REFL_A_RF9',
'LSC-REFL_A_RF45',
'LSC-ASAIR_A_RF45'
'LSC-X_TR_A_LF',
'LSC-Y_TR_A_LF',
'LSC-TR_X_QPD_B_SUM',
'LSC-TR_Y_QPD_B_SUM',
'LSC-POP_A_LF',
'LSC-REFL_A_LF',
'LSC-POPAIR_A_LF',
'LSC-REFLAIR_A_LF',
'LSC-ASAIR_A_LF',
'LSC-POPAIR_B_LF',
'LSC-REFLAIR_B_LF',
'LSC-ASAIR_B_LF'

The changes have been committed to the SVN.

Comments related to this report
travis.sadecki@LIGO.ORG - 15:45, Thursday 07 May 2015 (18312)

While looking over SDF Diffs after running the script, I noticed that the offset for ASAIR_B_RF18 (both I and Q) changed from ~0 to ~300 (a few order of magnitude), whereas the other PD offsets changed little.  Just a heads up.

evan.hall@LIGO.ORG - 20:38, Thursday 07 May 2015 (18320)

Sheila and I found that the dark offset for LSC-TR_X_QPD_B_SUM changed from −0.9 ct to −37.2 ct at 11:41:57 local this morning. Was this when the script was run? This value is way too big.