SRCL FF coupling remeasured
Sheila, Evan, Stefan

We remeasured the required SRCL FF filter today. The transfer function changed significantly.
The only thing we know we changed was the POP_45 phase (H1:LSC-POP_A_RF45_PHASE_R) from 66deg to 58deg - this minimized a SRCL drive in MICH.

The fit functions are roughly:

May 11: p:0,0  z:2.28571+31.9183i,2.28571-31.9183i,-180    (M1)
May 12: p:0,0  z:2.28571+47i,2.28571-47i,-600              (M2)

The 1st attached plot shows these two measurements, overplayed with the models M1 and M2.
The 2nd plot shows the 4 raw measurements (2 yesterday, 2 today).

We didn't try to engage this SRCL FF yet.
Images attached to this report
Comments related to this report
stefan.ballmer@LIGO.ORG - 04:35, Wednesday 13 May 2015 (18404)
At the end of the day - after sitting at 10W for a while, and with the MICH FF optimized - we measured the SRCL FF transfer function again. It looked a lot closer to yesterday's measurement.
gabriele.vajente@LIGO.ORG - 11:48, Wednesday 13 May 2015 (18409)DetChar, ISC

Those changes in the SRCL to DARM couplings are consistent with the simulation results for a changing SRCL offset. The reason of this changing offset can be related to many things, for example alignment.

Looking at the data and the simulation results, to have a zero at 30-50 Hz we need offsets of the order of 6-8 nm