Reports until 09:06, Thursday 21 May 2015
H1 CDS (SUS)
james.batch@LIGO.ORG - posted 09:06, Thursday 21 May 2015 (18552)
Shutdown and restart of h1sush2a
The h1sush2a computer was shutdown to allow a DAC card (D0) to be replaced in the I/O chassis, due to reports of a stuck voltage.  See alog 18543.

It was not possible to log in to this computer remotely, as it was not accepting new connections due to a "nf_conntrack: table full" error viewable on the console.  I was able to shut the computer down in an orderly manner from the console.  We have not encountered the nf_conntrack error in many months, but it appears when there's too many network connections being made.