The Y arm had not locked since the new 18bit DACs were installed. One problem was a bad DAC. DAC 0 was not giving linear outputs. I have replaced this DAC. The other problem was channel 0 on DAC4 which would not output any signal. After a reboot/IO chassis power down the problem with DAC4 was gone. We then had to replace DAC0 which seems to have fixed the problem.