Displaying report 1-1 of 1.
Reports until 14:54, Tuesday 26 May 2015
H1 SEI
thomas.shaffer@LIGO.ORG - posted 14:54, Tuesday 26 May 2015 (18629)
BRS Restart

Jim W., TJ

SYS_DIAG reported that BRS had flatlined, and after looking into it it looked like the software crashed on Saturday (attached image). The BRS is not currently being used for sensor correction so this was not a big deal but we did need to restart it. I took a trip down to EX, followed the instructions on Alog 13817, drove back and realized that that had not corrected the problem. I took Jim with me this time and he noticed that the rate box on the running analysis screen was empty. He restarted it again and it all seemed to work well. Everything looks good now.

Images attached to this report