Reports until 10:24, Thursday 28 May 2015
H1 CAL (CAL)
richard.savage@LIGO.ORG - posted 10:24, Thursday 28 May 2015 (18665)
Short Pcal test at Yend

DarkhanT, RickS

We are running a short (hour or two) test at Yend.

The Rx PD assembly is back in the Rx module.  The outer beam is blocked inside the Tx module so only the inner (upper) beam is reflecting from the ETM into the Rx PD assembly.

We are trying to identify the source of variations we have seen in the received power, but not in the transmitted power.