Reports until 13:24, Thursday 28 May 2015
H1 SUS
travis.sadecki@LIGO.ORG - posted 13:24, Thursday 28 May 2015 (18668)
SUS Driftmon updated

I updated the SUS Drift Monitor with values from a shortish lock stretch last night, 1116855522.  In addition, I updated the threshold updating script to widen the thresholds for PR2 and BS per Betsy's suggestion.  The IMs and OMs are still showing alarm, but we assume they will come back to green during a lock.  As part of the background to get the Drift Monitor back to functional, we (TJ at the keyboard and Betsy, Hugh, and myself in the peanut gallery) reassigned the alarm severity states (.HSV, .LSV, .LLSV, .HHSV) to their appropriate values (MINOR, MINOR, MAJOR, MAJOR respectively).  At sometime during the past month, these values were lost (i.e. NO_ALARM), presumably during some bootfest. 

We need to update the appropriate database such that these values are not lost during bootfests in the future.