Reports until 12:53, Monday 01 June 2015
H1 PSL
patrick.thomas@LIGO.ORG - posted 12:53, Monday 01 June 2015 (18751)
Ops PSL check
Laser is ON
Output power is 33.1 W (should be about 30 W)
Watchdog is active
PSL SYSSTAT: VB program online is red, LRA out of range is red (only VB program online should be red)

PMC
Last relock 6 days, 1 hour ago (should be days/weeks)
Reflected power is ~ 11% of transmitted power (should be 7% or less)

FSS
Locked 1 hour (should be days/weeks)
TPD is ~ 1.4 V

ISS
Diffracted power is around 5.5% (should be around 7%)
Last saturation event was ~ 1 hour ago (should be days or weeks)