Reports until 20:17, Wednesday 03 June 2015
H1 AOS (ISC)
evan.hall@LIGO.ORG - posted 20:17, Wednesday 03 June 2015 - last comment - 21:07, Wednesday 03 June 2015(18839)
SR3 oplev servo turned off

Sheila, Evan

The SR3 pitch oplev servo appears to be causing interferometer instability, so it is now turned off.

Starting this morning, we found that we could not lock the interferometer for more than 15 minutes or so, even at 3 W with DARM controlled with AS45Q. The symptom was that irregular spikes would start to appear in POP90 a few minutes after acquiring lock. One could also see the effect as 0.9 Hz ASC oscillation in the BS, SRM, SR2, dETM, and sometimes cETM loops.

We initially did not suspect the SR3 oplev servo, since it worked fine during last night's 10 hour lock. We instead tried turning off our ASC loops, but that did not solve the issue. We then sat in the DRMI configuration with arms off resonance and no ASC, but that did not solve the issue. We turned off ITM oplev damping servos, but that did not solve the issue. Then finally we turned off the SR3 pitch oplev servo, and we also disabled the top-stage PRM and SRM LSC offloading. This seems to have solved the issue.

Because Sheila and Elli had at one point today turned off the offloading but still were not able to achieve a long lock, we suspect this means that the SR3 oplev servo is at fault. However, it is not obvious to us why it takes a few minutes for this spiking to start happening (one would think that if it is caused by oplev glitching, then one would see the spikes immediately). Also, at first glance there does not seem to be glitching in the oplev sum.

Non-image files attached to this report
Comments related to this report
sheila.dwyer@LIGO.ORG - 21:07, Wednesday 03 June 2015 (18841)

It turns out that the problem we have been having today, which seems like a 1-5 Hz alignment problem, is actually caused by the verry low frequency offloading of PRM and SRM from M2 to M1.  The length feedbacks themselves are stable, and very slow, but there must be some interaction with the alignment that means that one or both of these make the IFO unstable.  

We're not going to try turning them on again.  We are turning the SR3 OpLev damping back on, since that seemed to solve the lockloss problems encountered on Travis' tuesday owl shift (alog 18777).