Our lock is at 14h and counting. Attached are the output control signals for the 10 optics that are controlled by the ASC system. (All of them are relieved to the top mass, so the top mass output is what I am plotting.)
There is some funny behaviour in PR2 pitch, and some of it also shows up in the BS pith nd yaw.
Also - interestingly, the sharp rise in SR2 PITCH about 10.5h ago (2015/06/06 07:55 UTC) corresponds to a drop in the inspiral range by about 10% (6Mpc), and comes from the 80Hz to 300Hz region. Whatever this noise is, it seems to be related to SRC alignment.
The third attached plot shows the difference between before that SR2 move (2015/6/6 7:32 UTC, 67Mpc) and after (2015/6/6 8:32 UTC, 61Mpc). Sounds like SRC ASC work is not done yet.