Displaying report 1-1 of 1.
Reports until 12:45, Monday 08 June 2015
H1 ISC
keita.kawabe@LIGO.ORG - posted 12:45, Monday 08 June 2015 (18989)
TMSX green clipping is not getting worse

https://alog.ligo-wa.caltech.edu/aLOG/index.php?callRep=16970

Since there's some chance that we're not going to go in EX in the upcoming vent, I checked the trend of X green QPD SUM over the past 5 months.

They only changed by 1% or so, and there's no reason to believe that it's getting worse. Certainly we will survive O1 without fixing it.

Images attached to this report
Displaying report 1-1 of 1.