Displaying report 1-1 of 1.
Reports until 11:57, Monday 15 June 2015
H1 SEI
hugh.radkins@LIGO.ORG - posted 11:57, Monday 15 June 2015 (19151)
EndX HEPI On HardStops at Rotation DOFs position. X Y Z not so much

Have locked HEPI at X End.

The rotational DOFs are within ~100nrads of the closed loop running position.  The Translational DOFs were let go as the Rots were dialed in:  X -5um, Y +67um, Z +108um.  These are the deltas from the nominally servo'd position.

Displaying report 1-1 of 1.