Particle counts taken during the removal of the HAM6 south & east doors on 06/17/15. Counts were taken with an older HPPC-6, using differential mode. HAM6 South Door Time 0.3µ 0.5µ 1.0µ Sample Location 13:24 100 0 100 In cleanroom – breaking bolts 13:32 300 100 100 In cleanroom – breaking bolts 13:35 0 100 0 In cleanroom – breaking bolts 13:43 40 20 20 Top of door flange venting purge air 13:48 3050 250 40 LVEA Room air HAM6 East Door Time 0.3µ 0.5µ 1.0µ Sample Location 14:07 2530 750 320 LVEA room air 14:09 960 350 320 In cleanroom – breaking bolts 14:30 0 0 0 In cleanroom – Door coming off 14:43 20 10 0 In cleanroom - Putting on the door covers 15:00 10 0 0 In chamber – Cover on 15:39 50 20 10 In chamber – Cover on