Laser Status: SysStat is good Front End power is 33.1W (should be around 30 W); 29.1W @ Power Monitor PD FRONTEND WATCH is GREEN HPO WATCH is RED PMC: It has been locked 2 day, 21 hr 15 minutes (should be days/weeks) Reflected power is 2.2 Watts and PowerSum = 25.7 Watts. (Reflected Power should be <= 10% of PowerSum) FSS: It has been locked for 2 d 21h and 29 min (should be days/weeks) TPD[V] = 1.49V (min 0.9V) ISS: The diffracted power is around 7.2% (should be 5-9%) Last saturation event was 2d 18 h and 5 m ago (should be days/weeks) NOTES: ISS diffracted power was up to ~9% this morning. I've adjusted the refsignal from -2.12 to -2.14 to yield ~ 7.5%