Displaying report 1-1 of 1.
Reports until 10:14, Wednesday 24 June 2015
H1 AOS
corey.gray@LIGO.ORG - posted 10:14, Wednesday 24 June 2015 (19304)
H1 PSL Check (My first attempt using TJ's new script)

Below is the output from TJ's script for running the "PSL Check" (No Issues were posted during this check):

Laser Status:
SysStat is good
Front End power is 33.04W (should be around 30 W)
Frontend Watch is GREEN
HPO Watch is RED

PMC:
It has been locked 0.0 days, 21.0 hr 28.0 minutes (should be days/weeks)
Reflected power is 2.287Watts and PowerSum = 25.64Watts.

FSS:
It has been locked for 0.0 days 21.0 h and 28.0 min (should be days/weeks)
TPD[V] = 1.477V (min 0.9V)

ISS:
The diffracted power is around 8.16% (should be 5-9%)
Last saturation event was 0.0 days 21.0 hours and 19.0 minutes ago (should be days/weeks)

Displaying report 1-1 of 1.