To test if NDS clients after tomorrow will correctly apply the upcoming leap second, I injected a sub-second signal into a DAQ DQ channel today. The H1:PEM-EX_ADC_0_08_OUT_DQ channel is a PEM test channel with ADC bit noise on its input. I ran awgstream to inject a signal which is one second long, and all zero except for 1/16th of a second starting at the 0.5S mark. This 1/16th second block is a step function, with an amplitude of 1000 counts.
The time of injection is chosen to be modulo 64 in GPS time, so it will be the first second of a 64second DAQ frame. The time of injection is
| GPS | 1119659968 |
| PDT | Jun 29 2015 17:39:12 |
| UTC | Jun 30 2015 00:39:12 |
The model runs at 16kHz, this channel is acquired to the DAQ at 2048Hz. The decimation filter is producing the ringing seen at the step transitions.