gerardo.moreno@LIGO.ORG - posted 17:03, Wednesday 01 July 2015 (19419)
Vacuum Activities @ Y-End

(Kyle, Gerardo)

Y-End station:
Connected and ran leak detector at turbo's exhaust in place of scroll pump, sprayed new, untested, conflats with 20 seconds of audible helium flow.
No leaks detected.
Turbo pump restored to original configuration.
Pump down continues.

Conflats tested: new Inficon gauge for ESD interlock, NEG pump 10" gate valve conflat only, Old TMDS valve + blank on middle port north door, New TMDS valve + elbow + blank on south door, and new blank for the removal of the old type gauge for ESD interlock.