Displaying report 1-1 of 1.
Reports until 09:53, Tuesday 07 July 2015
H1 PSL
edmond.merilh@LIGO.ORG - posted 09:53, Tuesday 07 July 2015 (19465)
PSL Status report
Laser Status: 
SysStat is good
Front End power is 32.8W (should be around 30 W)
FRONTEND WATCH is GREEN
HPO WATCH is RED

PMC:
It has been locked 13 day, 21 hr 9 minutes (should be days/weeks)
Reflected power is 2.5 Watts  and Power in Transmission= 22.6 Watts.
(Reflected Power should be <= 10% of Power in Transmission)

FSS:
It has been locked for 17 h and 34 min (should be days/weeks)
TPD[V] = .98V (min 0.9V)

ISS:
The diffracted power is around 6% (should be 5-9%)
Last saturation event was 2 h and 28 minutes ago (should be days/weeks)

NOTES: FSS TPD V is due for a tweaking.
Displaying report 1-1 of 1.