Nic, Sheila

We put an excitation into the HAM6 ISI ISO Y filter bank, (30000 counts at 0.3 Hz) from about 3:17-3:19 UTC July 11.  We then did a by eye fit  (on a log log scale)for a fringe wrapping model.  We expected the excitation to result in 30 um motion of the OMC, but we had to use 36 um to get the fringe speed right. We get an amplitude reflectivity of 1.6e-7 for the single pass shelf. (compare to 1e-5 measured in 17919)  We see no evidence of a second shelf or a shelf in the null stream.  

We plan to make measurements in exactly the same was as 17919, if we get a chance again tonight.

Images attached to this report
Non-image files attached to this report
Comments related to this report
sheila.dwyer@LIGO.ORG - 21:09, Monday 13 July 2015 (19609)

There is a typo in this alog, the reflectivity is r=160e-7, as the legend in the plot says, not 1.6e-7 as I wrote.