Nic, Sheila, Evan, Stefan - Turned on ITMX PIT oplev (ITMY PIT OPLEV was already on) - DC coupled SRM PIT oplev turned on. - Tried new SRCL input matrix YAW: AS 36 AI AQ BI BQ OLD: -0.6 0.3 0.3 0 NEW: 0 1.0 0 0 - This seemed better at higher power for RF offsets, but is too early for judgement. - Also tried new SRCL input matrix PIT: AS 36 AI AQ BI BQ OLD: 0 0 2 0 NEW: -0.5 -0.5 0 0 - We didn't keep this one though because it seemed worse at low power. - Attempts later tonight were hampered by winds up to 30MpH in the corner.