I wrote the script for the OPLEV charge measurements which sets most of the settings and easy-to-use. It works for ~ 2.5 hours. If you are the last person who leaving the LHO in the night - please, run it! Instructions: 1. Set ISC_Lock state to DOWN 2. Set both ETMs to "ALIGNED" state 3. Align Optical levers (pitch and yaw) for both arms to 0 +/-.5 urad 4. Run the scripts: scripts directory is /ligo/svncommon/SusSVN/sus/trunk/QUAD/Common/Scripts run the python files: ./ESD_Night_ETMX.py and the second script in another terminal: ./ESD_Night_ETMY.py If it works, it: a) In first ~ 30 seconds sets the channels and can warn about troubles with ESD or Alignment. If it happens - check this system and pressb) During the measurements it should once a second print "Receiving data for GPS second: 1234567" c) After all the measurements it should restore all the settings back. If it gives the errors but still receive some data - let it work. If it obviously not work, you can try to run it again. If it does not help - please, restore all settings running ./ESD Restore_Settings.py . Due to it is first try, today I will be very thankful if you'll check after running the scripts : 1. L3 Lock: Bias Voltage - 0 Offset - green light Ramp time - 5s 2. ESD linearization: Bypass - ON 3. For ETMY: turned on Hi-voltage driver. Just in case: scripts modify only this settings.