J. Batch, J. Kissel After failed attempts to run diagonalization measurements of the newest QUAD built using DTT (DTT opens, sends excitation out to the suspension as seen on MEDM, runs for the allotted time and stops, but no data is plotted and DTT reports an error message say "Test time out."), Jim found that the framebuilder / daqd process had "somehow lost 30 seconds ... possibly because we unplugged a cable temporarily." I think he's referring to timing cables that may have been plugged/unplugged during the Mass Storage Room gutting. He has rebooted the frontend, and I have burt restored using the ${RTCDSROOT}/userapps/release/sus/x1/burtfiles/x1susquad_safe.snap restore file. I'm retaking the diag. measurements now, and all is functioning properly.