Reports until 17:17, Friday 24 July 2015
H1 ISC (ISC)
stefan.ballmer@LIGO.ORG - posted 17:17, Friday 24 July 2015 (19919)
ARM offset now set in PREP_TR_CARM
Evan, Stefan

We kept having PREP_TR_CARM issues, and again tracked it down to moving dark offsets. now the guardian measures the offsets at the beginning of PREP_TR_CARM.

Let's see weather this finally eliminates those PREP_TR_CARM issues.