Reports until 21:47, Friday 24 July 2015
H1 ISC (ISC)
stefan.ballmer@LIGO.ORG - posted 21:47, Friday 24 July 2015 (19921)
SRCFF
Retook the SRCFF measurement. the data is in /ligo/home/controls/sballmer/20150724/FF
SrclFF20150724.xml & SRC_FFpath20150724.xml  (1st measurement)
SrclFF20150724v2.xml & SRC_FFpath20150724v2.xml (2nd measurement)

An initial fit produced a filter (FF4) that improved the performance at high frequency, but moo0.5 (the old moo, adjusted by a gain of 0.5) was still better at low frequencies. More fitting will be done during daytime. For now, guardian uses moo0.5.