Reports until 08:54, Tuesday 28 July 2015
H1 PSL
corey.gray@LIGO.ORG - posted 08:54, Tuesday 28 July 2015 - last comment - 10:58, Tuesday 28 July 2015(19985)
PSL Weekly Report

I forgot to run the the PSL Checklist during my shift yesterday.  Today, I ran our PSLweekly script and here is the output.  There was model work recently/currently done on the PSL today, so this is why many items are ~10min old.

Items to note:
ISS Diffracted power is HIGH!

Laser Status:
SysStat is good
Front End power is 32.51W (should be around 30 W)
Frontend Watch is GREEN
HPO Watch is RED

PMC:
It has been locked 0.0 days, 0.0 hr 10.0 minutes (should be days/weeks)
Reflected power is 2.061Watts and PowerSum = 24.37Watts.

FSS:
It has been locked for 0.0 days 0.0 h and 10.0 min (should be days/weeks)
TPD[V] = 1.521V (min 0.9V)

ISS:
The diffracted power is around 14.4% (should be 5-9%)
Last saturation event was 0.0 days 0.0 hours and 10.0 minutes ago (should be days/weeks)

Comments related to this report
edmond.merilh@LIGO.ORG - 10:58, Tuesday 28 July 2015 (19989)

Keep in mind that this was taken on a maintenance day after the mode cleaner had been taken down and a previous snapshot had been restored from prior to proper adjustment of the AOM Diffracted power. Not the most optimal time to be taking vital signs :)