Search criteria
Section: H2
Task: IOO
Alignment into HAM2
Locked JAC with RF at 1W. IMC WFS was centered in IOT2L and IMC started working. As we steered JM3 MC2 TRANS responded as expected (i.e. JM3 PIT -> MC2 YAW and vice versa though there's a significant cross coupling). Sheila had to recenter the WFS once again at some point as the WFS started to get off-centered as we turned JM3, but as far as the WFSs were centered, things worked.
In the end, MC2 TRANS was centered reasonably well (right before the cursor in the attached), JM3 DAC output was O(1e7) while a 28 bit DAC's range is +-134e6.
Apparently we made the last step of JM3 PIT adjustment in the wrong direction at the time (we didn't notice because the WFS is very slow) and overshot. I steered it back later, now H1:SUS-JM3_M1_OPTICALIGN_P_OFFSET is -39 instead of -19 (second screen shot).
Anyway, this means that the alignment into HAM2 is good with more than a comfortable range left for JM3. No need to further refine.
One caveat is that the IM4 trans is totally off in PIT. But that's a downstream problem which we'll have to deal with later in HAM2. That should not prevent us from moving forward to close down the chamber.
The tasks listed below are described in different alogs.
JAC TRANS PD calibration
Last beam dump (alog 89249)
ALS beam path check
Summary: JAC TRANS PD well aligned, rough power budget done, ALS beam from PSL does not hit any wires or components before reaching its intended steering mirror in HAM1.
POWER BUDGET
Jason and I went into do some measurements on the TRANS PD path after 11am. We were unsure yesterday that this PD was well aligned after the laser window installation.
The beam going to the TRANS PD matches roughly what we expect with the uncoated laser window we now have as BS1. Here is a photo showing the power Keita and Jason were getting after swapping in the laser from yesterday (in the lower left plot).
There was approx 0.03 on TRANS PD. Previously with a HR mirror in place of BS1 we had ~ 3 on TRANS PD (see lower left plot on ndscope).
Keita did a calculation of the rough power reflected from BS1, assuming an AOI of both 40 degrees or 35 degrees (AOI should be around 39 degrees according to layout).
rp=tan(theta-phi)/tan(theta+phi)
theta is the AOI and phi is the angle of refraction,
phi = asin(sin(theta)/n),
n=1.4496 for fused silica at 1064nm.
rp(theta=40deg)^2 = 1.14%
rp(theta=35deg)^2 = 1.63%
This means that the beam is likely not clipping on the TRANS PD as the power on it seems to have scaled as we expect with the laser window installation.
To double check this and provide a calibration for this PD we did some power meter measurements in chamber.
We measured the output from the input side curved mirror of the JAC (Te2), with the PSL set to 1W output using the rotation stage (otherwise it is hard to see this beam on the card).
JAC Te2 = 2.7 mW
JACT_BS1 transmitted beam = 2.4 mW
JAC TRANS PD beam = 37 microW
This leaves us with 0.263 mW unaccounted for in transmission, this puzzles me.
We put the power out of the PSL down to 100 mW to measure the input power and output power in HAM1.
JAC input power = 115 mW
HAM2 input power/after HAM1 output periscope= 96mW
ALS BEAM CHECK
Jason opened the light pipe and checked that beam does not intersect anything int he new installed path before reaching its SM that directs it towards ISCT1.
We installed JM1, balanced it and aligned it. A beam dump was placed behind it though we could not see the transmission with 1W input.
After this, JAC locked with RF without any problem though the input was wobbly when the purge was up.
We searched for unexpected ghost beams (also with 1W input) and didn't find any.
We uninstalled many (but not all) temporary dog clamps and irises.
We revisited the IMC alignment because it's been off in PIT since Thursday or Friday. We locked JAC using dither (because we wanted to turn down the purge air). We enabled the IMC WFS just for MC optics and steered JM3, but weren't able to center the MC2 trans. Steering JM3 just made the IMC transmission worse while making not much impact on the desired degree of freedon (JM3 PIT -> MC2 trans YAW, JAM3 YAW -> MC2 trans PIT).
Tomorrow, we'll revisit the IMC alignment. We'll also measure the power coming into JAC TRANS PD as well as the actual transmission of JAC while locking it with RF so we can use JAC trans PD as the measure of the power into HAM1.
Jennie W, Keita K, Rahul K, Ryan S,
This morning Keita turned the power up to 1 W and the purge air down in order to scan the OMC and measure the modulation index.
We had some work to do to lock the JAC, the cavity wouldn't scan properly when managed by the guardian code.
Ryan and I have looked back at it and think there may be something wrong with the feedback to the PZT. It can't be the PZT itself though, because Keita and I were able to lock on the TRANS PD signal with the dither locking done manually by tuning the PZT offset to find the resonance and then turning on the feedback.
I have enclosed a picture to aid problem solving next week.
We then had problems keeping the IMC locked so Keita had to change the threshold used to check if the cavity is locked down to 2. This was changed at line 133 in ISC_library.py at userapps/h1/isc/guardian/.
trans_pd_lock_threshold = 2
We then managed to do three scans,
one with 45 MHz off, one with both on, and one with 9 MHz off. Detailed analysis in progress.
After turning the power back down Keita and Rahul swapped the mirror we currently have in front of the TRANS PD, with the uncoated laser window that is meant to be there (JACT_BS1).
They had to top gun and swab it to remove two dust particles first.
This replacement means the power to the PD is lower and so we needed to power up to 1W to check this alignment.
I have left the mirror we took out on the table on the -Y side labelled as 'HR mirror taken out of JACT-BS1'. Need to check which exact mirror this is with Masayuki on Monday.
Keita locked the cavity manually with the REFL PD and optimised the amount of light falling onto the TRANS PD.
I powered down to 100mW and de-energised the rotation stage and closed the light pipe at the end of the day. We also had the rotation stage de-energised and locked out when we were working in chamber.
FSR = 265 MHz (see this ref and table on P.110).
FSR in s on scan = 8.39 s
For the 9 MHz on the measurement with both sb at nominal power: : 0.0103mA is the upper 9 MHz sb vs. 0.745 mA carrier (left-hand TM00 peak).
m_9 = 2*sqrt(0.0103/0.745) = 0.235
For 45 MHZ in same measurementdata : 0.0151 mA is the upper 45MHz carrier vs. 0.745 mA carrier (left-hand TM00 peak).
m_45 = 2*sqrt(0.0151/0.745) = 0.285
The 9MHz seems similar to this (alog #89001) measurement with the previous crystal (0.26) but the 45MHz is different (0.31).
Modulation index measurement for 45M and 9M, they're healthy.
1W into JAC, no WFS for IMC, ITMY single bounce. Scanned OMC with both 45MHz minimal power (~4dBm) and 9MHz full (~27dBm), then both in full power, then only 45MHz in full (see screen shot).
Jennie will post the modulation index later, but the peaks look healthy to me.
The alignment into IMC was rather off in PIT and the guardian had a hard time locking IMC (I changed the lock threshold in the guardian). We didn't move MC, didn't bother to do any thorough investigation, but either JAC or MC mirrors moved. Maybe we have to lock IMC, use WFS for MC, and steer JM3 and see if the MC mirrors will be driven far from where they are. If they are we might have to touch up in-chamber alignment again.
JAC TRANS BS was swapped with the real one (laser window with AR only on one surface).
We replaced the JAC TRANS BS (which was a temporary high reflector until today) with the real one that is just a bare glass (the reflectivity is 7% or so depending on AOI).
We locked JAC with RF, was able to see the beam coming to the TRANS PD but wasn't able to see the reflection of PD, so we raised the power to 1W.
Initially the beam was missing the PD because the new optic is thinner than the temporary one. We realigned the optic to steer the beam back to PD. Since we could not directly see the PD surface, we just steered the beam up and down, left and right until the PD output starts to drop, and put the beam roughly in the middle.
We were able to see the PD reflection using an IR card and an IR viewer. There were two reflection blobs (which was the case before, too), and I moved the JAC TRANS BD in +X direction by 1/4~3/8" to catch both.
See results here (alolg #89230).
Beam dump for HAM1-HAM2 septum window AR reflection.
This was installed.
Beam dump for the -Y door viewport reflection.
Jason and Betsy installed it. There seems to be a discussion as to whether or not something else needs to be done.
EOM and JAC power budget again.
We measured the power at various places. While Jason held the power meter head still, I ran the statistics function of the power meter for a few seconds. I only list the mean and the standard deviation.
Wrong pol is king of large, ~0.2% of the main beam power coming out of EOM. That's 200mW when 100W goes through the EOM.
JAC throughput of 92% is not great, but Jason says the alignment and the matching are not really optimised.
| JAC input | 112mW+-462uW | |
| EOM input = JAC output | 103mW+-6.3mW |
JAC OUT/IN = 0.92+-0.06 |
| EOM front AR reflection | 57uW+-4.4uW |
EOM AR/IN = (5.5+0.5)*1e-4 |
| EOM output (including the wrong pol beam) | 96.9mW+-1.4mW | |
| EOM wrong pol beam | 226uW+-17uW |
EOM wrong pol/IN = (2.2+-0.2)*1e-3 |
| EOM main pol (= out total - wrong pol) | 96.7mW +-1.4mW |
EOM main pol/IN = .9978+-2e-4 |
Afternoon work, REFL path aligned, RF lock works (Jennie, Jason, Betsy, Daniel, Keita)
It seemed as if what was supposed to be TFP after HWP was not really TFP, it was temporarily set aside.
REFL path to the JAC REFL RFPD was aligned without TFP. 100mW into JAC was enough to see the REFL beam there.
DC responded as expected.
There was a confusion about which demod chassis was used for JAC, which was sorted out by Daniel who subsequently set the demod phase. I zero-ed the dark offset.
I copied the JAC lock filter from dither path to JAC-L_SERVO path, locked JAC with dither, disabled the input to the dither servo and enabled RF locking in parallel, which worked just fine. I made rough changes to the RF servo to bump up the UGF to ~400Hz without too much gain peaking, I haven't tried anything aggressive to squash the residual motion below 200Hz, you might want to tweak it further.
I didn't disable the dither itself for the JAC PZT so we can compare the spectrum of RF and dither side by side. See attached, this was measured with 100mW into JAC, note that the dither error signal is scaled. References are with the purge air on and current traces are with the purge air completely turned down.
Guardian needs to be changed to allow smooth locking with RF.
I added this shutter (shutter H1:SYS-MOTION_C_SHUTTER_M) to the main shutters screen at sitemap->LSC->Shutters. I also added a menu button that takes you to the control screen to the sitemap->IOO->JAC Overview screen, see pic.
We locked JAC with 1W input to find ghost beams. Details will come after we're done with the septum window reflection, but anyway here is the list:
None of the new beam dumps are interfering with the main beam and JAC refl/trans.
Jennie W, Ryan S
Summary: Ryan and I updated the JAC and IMC ASC models. We installed JM3 yesterday and so now the IMC cannot use the PSL PZT mirror as an alignment actuator. I have committed the changes to the svn but erik and dave will do some checks tomorrow before they commit to the revision locked version of the model and restart the DAQ. So changes are not 'live' yet.
The edits were made to h1ascimc.mdl which has top level blocks for IMC and JAC.
The JAC top level model sends signals directly to the PSL PZT mirror and these degrees of freedom are swapped as the PSL PZT basis P and Y are switched before input to the JAC due to the HAM1 input periscope.
I took out the feedback paths for PZT_P and PZT_Y that come out of the IMC block. The picture shows the old config and I have highlighted what I removed.
At the top level, IMC no longer sends signals to the PZT but instead to JM3. On the top level diagram, the signal JM3_P is sent via PCI cards to the channel "H1ASC-JM3_YAW_SUSHTTS" as pitch in the JAC basis is yaw in the IMC basis due to the HAM1 output periscope. The picture shows the old config and I have highlighted what I swapped.
The signal JM3_Y is sent via PCI cards to the channel "H1ASC-JM3_PIT_SUSHTTS" as yaw in the JAC basis is pitch in the IMC basis due to the HAM1 output periscope.
Within the IMC top names block I removed the output channels for the PZT from the WFS feedback path, we already had paths to feedback to JM3 within this path. The picture shows the new config.
I removed the PZT locking path from the LCKIN block and replaced it with one for JM3. The picture shows the new config. I am not sure we ever use this path (which is for dither asc control of the IMC) in the first place so maybe this was unneccessary.
I also removed the channels H1:IMC-PZT_YAW_OUT and H1:IMC-PZT_PIT_OUT from the DAQ channel list for the IMC model.
Jennie W, Dave B, EJ D, Erik V, Keita K, Daniel S, Jeff K, Olli P,
This morning I rebuilt the model again before it was restarted to add back in the IMC-PZT_PIT_OUT_DQ and H1:IMC-PZT_YAW_OUT_DQ channels in the top level of h1ascimc so as to avoid removing these channels from the GDS broadcast channel list.
Unfortunately this change did not get propagated to the live model today (it was not built with the rev-locked tag). This caused some confusion when the model + DAQ restart happened ( alog #89196). I had forgotten to SDF the PZT sliders for IMC-PZT_OUT and so this mis-aligned the beam to the JAC. Keita brought this back by altering the IMC sliders and we were able to continue with optics work in HAM1.
On the CDS side, none of my changes from today or yesterday have been uploaded to the 'rev-locked version of the model so everything was restarted with the old model config from yesterday morning. We will aim to restart with model changes to h1ascimc (that I made last night and this morning) at sometime Monday morning.
We refined the alignment from JAC to IMC by iterating small amount between JM2 and JM3. It was hard due to purge air and the suspended JM3, instead of relying on the IMC scan and minimize 1st order modes, we locked IMC and minimized the IMC REFL DC on average. After that the beam was still centered on JM3 well.
We closed the chamber and turned down the purge all the way and the alignment was indeed good. There was almost no 01 (PIT) mode power, 10 (YAW) mode power was less than 1% of 00 mode power, and the 20/02 mode power was 0.23 to 0.24%. See MM.png.
(To identify which mode is what, I intentionally misaligned JM3 in YAW (that causes PIT misalignment for IMC. See mode_identification.png.)
Or, rather, a better JM3 integration and PSL unintegration.
I made a temporary IMC_WFS_MASTER and IMC_WFS_OUTMATRIX_kk screen such that it's easy to route IMC WFS signal to JM3, not the PSL PZT, because I wanted something that works now.
However, right after I made what looks to be an OK screen, I realized that this doesn't work. PIT signal should be routed to JM3 YAW and vice versa. No fully-working IMC WFS until the next model update.
In addition, earlier today Daniel suggested to nuke PSL PZT from the IMC ASC (good idea).
If you want to revert back to the old medm, copy the backup
/opt/rtcds/userapps/trunk/asc/common/medm/imc/IMC_WFS_MASTER_BAK_20260218.adl
to
/opt/rtcds/userapps/trunk/asc/common/medm/imc/IMC_WFS_MASTER.adl
This morning we have pushed the JAC EOM by about 0.6mm (using ~25 thou thick washers) in -Y direction, following the finding of last Friday (alog 89158).
After that the beam was good on the input side plate (the beam is offset in +Y direction by 0.1mm) and was OK on the output side plate (0.5mm offset in -Y direction).
The beam position on the crystal itself should be ~0.13mm in +Y direction on the input face and ~0.36mm in +Y direction on the output face. The angle between the nominal path and the actual path outside of the crystal is about 0.6 degrees. See pictures and cartoon.
Calculation depends on the refractive index, I assumed n~1+deflection/wedge=1+2.35/2.85~1.85, but using 1.85+-0.5 instead won't change anything in a meaningful manner.
This is acceptable, the beam is more than 1.5mm away from the side face of the crystal, cannot remember the beam radius but it should be smaller than 600um if FDR is still valid, so it might be 2.5 beam radius or maybe more.
IFO REFL beam check was done.
After Jennie restored the IMC alignment to post-IMC axes check state, IMC was locked, PRM was alignmed and the IFO refl beam in HAM1 was quickly checked to see if the REFL air path somehow interferes with the new POP periscope stiffener. It didn't.
JM3 swap is ongoing.
Partly in the interest of time, I asked others to go ahead. Rahul and the team are working on it right now.
Yet to be done items:
I calculated the mode-matching before we replaced JM3 and got a limit of 0.26 % for the mode-mismatch as the TM20 mode was hidden in the noise at 100mW input power. We turned up the whitening gain to 42 dB from 30dB to have a better chance and still couldn't see it.
This plot shows the zoomed out ndscope of the TM00 modes and this one shows the max value for TM20.
After JM3 was installed and its position, pitch and yaw had been tuned by Rahul and Betsy to optimise the pointing through our HAM1 irises, Keita, Jenne and I tried to tweak up the alignment with JM3 sliders.
I have left the sliders near here and could not get them much better.
I measure the mode-matching to be 0.43 % with this alignment which is worse by at least a factor of two.
See photo of TM20 mode here.
The 10 and 01 modes are much higher than they were previously, so we will need to do some alignment of the fixed JM2 or JAC_M3 mirrors.
Note for the MM measurement we were accidentally scanning with the MC2 length and the PSL laser frequency so this might make the measurement confusing.
I closed the light pipe and turned up the purge air before going home.
Let me point out that the term “mode matching” used in Jennie’s post is not exactly accurate in this case; it would be more precise to refer to it as the TEM20/02 mode peak fraction. Since there is a large misalignment, the second-order modes are also enhanced. Therefore, that contribution should be subtracted before attributing the remaining fraction to mode mismatch.
Jennie W, Ryan S, Keita K
Summary: MC suspensions back to good time from 10th of Feb, we should only lock the IMC length loop manually during this installation period to avoid the WFS engaging.
Today Ryan and I had the task of figuring out why the IMC suspensions seemed to be badly aligned on Friday.
We found a good reference time 16:48:29 UTC on Tuesday 10th Feb at the end of the period where Olli, Jenne and I set the IMC axis back to its nominal state before the vent using ASC loops for DOF 1 and 2 of the IMC plus manual moves of the JM3 mirror in chamber as this cannot be moved by an ASC loop as it is not a tip-tilt currently. See photo of the WFS and MC mirror top mass OSEMS here. Note it is important to choose a time when IMC is in the locked state as the MC2 mirror gets mis-aligned when the IMC guardian is offline.
We figured out that the alignment was changed about 19:33:17 UTC on 13th Feb probably by the IMC guardian as Elenna used it to lock the IMC for phase check measurements. See photo, where the vertical cursors are set at the good reference time and today.
After some investigation of the guardian we reminded ourselves that DOFs 3-5 are turned off in the IMC WFS MASTER filter banks. DOF4 and 5 are normally off and DOF3 is intentionally left off because of the replacement of PSL PZT actuator by JM3.
When the IMC guardian is used to lock it automatically triggers the WFS through some logic from the simulink model, and so when Elenna used the guardian to lock it turned on the feedback to DOFs 1 and 2.
Until we install the JM3 tip-tilt and are finsihed most commissioning of the HAM1 hardware we should only lock the IMC manually by engaging the MC-L servo using the common mode servo board controls.
Keita and I zeroed the M1 LOCK fiters for all 3 MC mirrors and I changed the alignment sliders so the M1 osems were back to their values from the reference time on the 10th. See photo of ndscope.
To ensure the IMC WFS loops don't cause the MC mirrors or the PZT upstream of JAC to move on us like they did last week, we've turned the outputs of the IMC-DOF_{1,2,3}_{P,Y} filters OFF. Since these outputs are not touched by the IMC_LOCK Guardian, we can now safely use the Guardian to lock the IMC without the concern of WFS turning on and changing alignments.
At some point, the h1ascimc model will need to be updated to send the output of the DOF_3 loop to JM3 instead of the PZT (and medm screens updated too), as that's what's immediately upstream of the IMC now.
Jennie W, Jason O, Keita K, Betsy W, Camilla C
Summary: EOM alignment and position captured, moved to opics lab for crystal replacment, JAC-REFL path alignment in progress.
This morning Jason, Keita and I checked the centering of the HAM1 irises after the JM3 alignment Olli and I did on Tuesday. We didn't move the irises right after JM3 as this seemed centred. We did reposition the iris right before the HAM2 septum plate in x and y.
This, along with the alignment of the IMC Olli, Jenne and I did on Tuesday, captures where the beam was with the current EOM installed.
Keita marked the EOM base position with a dog clamp and removed it to the optics lab for crystal replacement. The shims for the EOM and the screws that hold it to the table are in a foil packet on the +Y - X HEPI pier.
Betsy, Camilla and I rechecked the REFL path and spent some time moving JACR-M2 to centre it on the incoming and outgoing beams. Then we pitched down this mirror so the outgoing beam hit the upper periscope mirror. After this Betsy had some concerns that we were too close to where the JM3 tip-tilt will sit (we are currently using a siskiyou mount for this mirror). So we yawed the JACR-M2 mirror so the beam hits the viewport slighly more to the right (as viewed from -Y side). Looking at the placement on the viewport simulator, the beam is low and right on the viewport, whereas it comes to the table, high and right in the bellows hole. This table it almost at the top of its adjustment range, so Jason and I plan to go in this afternoon and see it we can move up the table feet so the beam can move up in the viewport a bit. Betsy and I measured the tilt of the beaM from JACR_M2 off the table. The heigh is 103 mm at the table hole 8 inches from the -X edge and 8 inches from the +Y edge. It tilts up to 113mm at the last hole in this row on the -Y side.
Betsy and I also checked the pitch of the beam entering and leaving JAC after the EOM was removed. The input beam to the JAC was 102 mm above the table, the output beam before the L1 lens was the same height, and the beam height just before JAC_M3 was also 102mm. There is not a large distance between the JAC and JAC_M3 steering mirror so it is hard to get a good measurement of any small pitch in the beam. Pictures for this alog are coming.
EOM crystal was swapped, EOM was tuned in the lab and put back in place in HAM1. Tuning is good. First look at the beam looks good, no ghosts.
We extracted the EOM from HAM1 after the position of the EOM base was marked with three dog clamps.
In the lab, the EOM pivot plate was separated from the EOM base (which was a major pain again), then the EOM top structure was separated from the pivot plate.
Before unmounting the old crystal, three set scews in the face plate were screwed in to contact the electrode board, and then very slightly backed off. After this, we lifted the board/side assy from the front plate to expose the crystal on top of the face plate, swapped the crystal (in our setup, the distance between the crystal edge and the front plate edge close to the output side was set to ~7.5mm due to 0.5mm shim washer we use between the front plate and the input side plate), and put the board/side assy on. This was much easier than before due to the aforementioned set screws. Elenna will post some pictures.
In the afternoon the EOM was tuned in the lab and put back into HAM1. We didn't bother to tap things around this time.
EOM was transferred to HAM1. Tuning measurement was repeated in chamber, no big change from the lab measurement and they're good (see pictures).
With EOM in place, we locked JAC and confirmed that the wedge orientation of the crystal was correct (because the beam was mostly deflected in YAW toward +Y direction). We also saw that the beam deflection was different from the old EOM in PIT as well as YAW (more difference in PIT according to Jason).
No fine alignment of the EOM was done for today, but we quickly raised the power to 1W and neither Jason nor I were able to find any clear ghost beams, unlike with the old crystal.
Below is a table of RTP crystals at LHO (see alog 89125, alog 89115).
| RTP crystal S/N, batch | Status |
| 10252007, Old | Good, in HAM1 |
| 10252003, Old | Chipped. Probably never used. |
| B1913109, New | Uninstalled, ghost beams |
| B1913108, New | Never used, ghost beams |
Here are some further notes and photos of the replacement work:
Following alog 89115, we found that the old batch crystal from that alog (S/N10252003) had a big chip at one corner. It is pretty bad we don't want to use that.
Betsy found another old batch (S/N10252007, "inspected 12/21/11" and UF tag dated 4/21/09), so we A-B-ed that one with the spare new batch (S/N B1913108).
The beam path was made as level as possible at 3" height using a beam leveling tool (a black metal thing with a tiny aperture at each inch of height).
We put the crystal on a platform that is roughly 2" 29/32 (which is about 2.4mm lower than 3"). The crystal is 4x4x40mm so that's about the right height.
We spent some time to make YAW alignment as good as we can for each of the crystals.
We scanned the beam in PIT from top to bottom (or bottom to top), each extreme is where the beam is almost clipped (but not actually clipped) by the top or the bottom face of the crystal.
Look at the attached, the new batch (left column) clearly shows multiple beams even though the focus is not as sharp as the old batch photos. As we misalign in PIT, the dark place moves relative to the main beam and the contrast changes too, but multiple ghost never went away. At the extrema (very close to the top or bottom edge) it looked as if the beam is better but I'm not sure it actually was.
The old batch (right) didn't show such a behavior. The beam shows something like a diffraction pattern but no separate ghost beams. Everything moved with the main beam. Not sure if the diffraction pattern came from the aluminum surface or EOM, but clearly this is MUCH better than the new batch.
Note, due to the apparatus (the steering mirror is 20" upstream of the EOM), we haven't searched in a huge PIT angle space, it's actually roughly +-4mrad or so, the angle is not negligible but it's more parallel displacement scan than an angle scan.
Also note, when the crystal was put in place it seems that there's some vertical deflection which was different for the old and the new. On the top two pictures, there's no change in the input alignment into the crystal.
Based on this observation, I'd say using the old batch makes sense. LHO people (Jennie, Rahul, Betsy and myself) had a brief conversation with Masayuki and MichaelL and we all agreed that that's the way to go.
Attached is the picture of the chip on the spare "older" crystal S/N10252003 The other picture shows the box labels of the EOM crystals and stat at LHO, namely: 10252003 chipped 10252007 to be swapped into the JAC EOM 2 newer ones which are having some scatter issues as Keita has written about
It took much longer than expected but we set up the beam path for the RTP test in the OSB optics lab.
Since more power makes it easier to see the ghost beams, I removed the beam dump that used to receive most of the red power (~530mW) and directed the beam to the front of the table (red path in the attached). I stole the steering mirror that used to be used for the low power P-pol path (circled in red). The low power p-pol path is now simply blocked. No other change was made to low power S-pol path (orange) as well as green path (green), but the beams are blocked by beam dumps. If you want to use these, simply unblock.
The beam radius will be 300~400 um or so at the location we plan to put the RTP (represented by a green rectangle in the second attachment). Elenna will post the plot of the beam size measurement.
The third picture shows the containers for different RTP. Left is the one for the crystal in HAM1. The middle seems to be from the same batch. Right looks different, on the bottom of the container there's a label saying "I/O something something 2017" so this is likely the old one.
We didn't have time to actually test the crystals, wait for tomorrow's udpate.
I made a mistake when providing calculations to Keita about the beam profile- I incorrectly input our distances as mm instead of cm. However, I think it's ok overall.
Keita and I put an available lens (f = 286.5 mm) into the beam path, and then used a thorlabs profiler on a rail to profile the resulting beam at five points. We measured distances from the lens to the profiler and accounted for the set back of our profiler from the edge of the mount, etc. This measurement allowed us to measure that the beam waist is roughly around the location of the laser, and is about 130 um in the x direction and 202 um in the y direction. Unfortunately, the beam quality isn't great, this is the best we could do. (Note, because of my mistake we chose not to use this particular lens, but it probably would have been fine for our measurements after all).
After some iteration, we determined that a f=401 mm lens was suitable, and we ended up placing it pretty close to the original lens location. We ran another profile measurement and found that we could achieve a beamsize of about 313 um in the y direction and 251 um in the x direction (different than Keita's reported numbers above because I originally fit an incorrect seed waist).
I have attached two plots. The first shows the profile of the beam with the original lens, and the second with the resulting lens that we have now used to measure the EOM crystal.
So, the beam is maybe a bit smaller than the beamsize on HAM1 that goes into the EOM crystal (around 350 um).
Jennie W, Jason O, Keita K.
As reported in this alog (#89073) from Masayuki and Keita, after we turned the power in HAM1 up to 1W we found a series of vertically spread ghost beams aroubnd the main beam after the EOM and before JM3.
These could not be removed by translating, yawing or pitching the EOM position relative to the beam. It was decided in a larger meeting with EOM design personnel that we would first check if the crystal was cracked or damaged anywhere in case this is the cause.
First photo shows the EOM from above, using a green torch to illuminate the beam path. I can't see any scatter from defects or cracks in the crystal.
Second photo shows possibly a chip at the corner, but this should not affect the beam as its right at the edge.
Third and fourth show side view with illumination from the top at an angle.
In summary we did not see any 'smoking gun' to cause these ghost beams.
Very rough power estimate for the ghost beam(s) is ~O(1%)
Jennie and Jason set up another temporary iris between JM2 and JM3, centered it with 1W into HAM1 to carefully block the ghost beams without blocking the main beam, then changed the power to 100mW (for safety) and measured the power at various places. Measurement accuracy cannot be great (Jennie and Jason says the numbers were jumping around as it was difficult to hold the power meter head at a fixed position mid-air) but I would say the power in the ghost beams is ~O(1%).
| JAC out | ~105mW |
| Between JM2 and the iris (includes wrong-pol beam) | 104~105mW |
| After the iris (wrong pol as well as ghosts blocked) | 99~100mW |
| Wrong-pol beam | 1~3mW |
| Background light (no beam) | 1~2uW |
Where do they go?
After opening the temporary iris that we just put in all the way, the iris just downstream of JM3 was already blocking some of the ghost beams as well as the wrong polarization beam (JM3iris.jpg). Vertical beams don't look vertical because the iris is not a flat plane and we have a large parallax here. Anyway, it seems that we can block further if we want to from the top and the bottom.
The picture of the last iris on HAM1 shows that something is blocked on the left (+Y) side (outputiris.jpg). Looks like the iris is clipping something on the right but the camera couldn't be positioned to have a good view for both sides.
The last picture (after_last_iris.jpg) shows the beam right after the last iris on HAM1. You can see that some ghost beams are still coming through.
With this beam injected into HAM2 and misaligning MC2, we looked into IOT2L to see the MC REFL beam. We weren't able to find ghost beams there, though Jason and I felt that the beam is not super clean.
One question Jason had was whether or not the diverging beams that originate from the EOM location are supposed to keep diverging after lenses.
The beam after the second lens is actually not diverging. According to this plot, we suppose to be able to find the splitted beams in the IOT2 table.
EPO taggin'.
Summary of investigation into the vertically split beam from the EOM
First, we confirmed that the vertical beam splitting observed yesterday originates from the EOM itself. To check the possibility of multiple reflections from lenses, we inspected the back-reflection port of the JAC output mirror. Two reflected beams were observed, most likely originating from the planar and curved surfaces of the lens, and they were mainly separated in yaw. Since the space between the EOM and the lens was blocked during this test, these reflections were conclusively identified as lens reflections. No vertically split beams were observed from this source.
A knife-edge–like test was performed by slowly lowering a metal ruler from the top at both the EOM input and output. At the EOM input, the entire beam disappeared simultaneously, whereas at the output the beam disappeared gradually from the top. This behavior confirmed that the vertical splitting is generated inside the EOM.
To accurately determine the beam positions at the EOM input and output, beam positions were measured from photographs. Taking refraction and geometry into account, it was found that the beam is slightly displaced in the horizontal direction. Details of this analysis will be documented in Keita’s alog.
Based on this result, the EOM was rotated in yaw. Dog clamps were placed at the ±y projection points of the EOM input and output, in contact with the base plate. One 0.5-mm shim was inserted between the base plate and each dog clamp to rotate the EOM counterclockwise. However, the beam pattern did not change. Additional shims were tested, but no significant change was observed. The EOM was fixed with one 0.5-mm shim at each position.
Next, pitch adjustments were explored. The original shim configuration (two shims on the +y side and one on the −y side) was changed by moving shims to the +x side (one location at center) and the −x side (two locations, upper and lower). Each location initially had two shims, and by adding or removing shims, it was observed that the vertical positions of the split beams changed. When the +x (downstream) side was raised, the vertically split beams appeared in the upper part of the beam profile (approximately 5–7 beams; Keita will upload photos). Conversely, when shims were added to raise the −x side, the split beams moved toward the lower part of the beam. With the downstream side lowered by approximately 0.25 mm, about two beams were observed in the upper part and one in the lower part.
1W input power shows 6 or more beams, but 2-3 beams can be observed even with 100mW.
With the last configuration, we proceeded the IMC scan measurement after alignment. The 2nd order mode peak was the same level as we observed when we sim up the EOM first time.
In summary, the EOM shows highly questionable and nontrivial behavior. Possible causes include diffraction due to crystal defects or multiple reflections at the AR-coated surfaces. However, identifying the exact mechanism is challenging at this stage.
Horizontal beam position offset on the EOM input and output aperture on the side plates.
We realized that the nominal beam position on the EOM input and output aperture is NOT centered on the crystal cross section projected onto the side plate face, the beam is horizontally offset in +Y direction.
Look at the first cartoon (cartoon.jpg) and references therein. The beam spot offsets are 0.91mm on the input side plate and 0.54mm on the output side plate, respectively, assuming that the beam deflection angle per surface of EOM is 2.35 degrees as implied in D2500130.
0.91mm is not a small offset, it's almost 1/4 of the crystal thickness (it's 4x4x40mm).
This means that the beam should be (see nominal_sideplate.png, note that the drawing scale of the input aperture in this is twice that of the output aperture):
~3.9mm from the left (+Y) edge of the visual alignment aid notch on the input side plate,
~3.2mm from the right (again +Y) edge of the aperture hole on the output side plate.
Measurements, adjustments and measurements made the beam closer to the nominal location.
Based on the above knowledge, we took pictures of the beam position on the input/output aperture, paying attention to the errors that could arise from the parallax (which is unavoidable), i.e. the sensor card should be as close to the face of the side plate as possible and the beam spot on the sensor card should be as close to the sentor of the camera sensor as possible. This was a tougher job than you think.
Anyway, in the first round of measurements, we convinced ourselves that the beam was:
off in -Y direction by 0.7mm relative to the nominal beam position on the input plate of the EOM,
off in +Y direction by 0.5mm on the output,
give or take 0.2mm or so (the error is based on two pictures for the input beam position with random variation in parallax coming from camera position and the distance between the side plate surface and the viewer card).
We rotated the entire EOM base by using two dog clamps against the EOM base and inserting appropriate shims (EOM_rotation.png). We didn't use the YAW adjustment feature for the EOM pivot plate because there's no way to rotate it in a controlled manner.
After the first adjustment we thought that the beam coming out of the EOM looked better (which might have been false). On the second adjustment the beam looked the same or slightly worse (which might have been false) and we reverted back to the same position as the first adjustment.
Multiple beams mostly in PIT coming out of EOM (pictures and history)
1W into HAM1, otherwise it's hard to photograph these clearly.
The first picture is right after the YAW adjustment was made but before adjusting PIT. The card is held just ABOVE the main beam, you can see four blobs that look like some kind of ghost beams. (If you try to picture the main beam, it's so bright these ghosts become hard to capture.)
The second picture is after the first PIT adjustment. You can only see maybe two blobs, but later we found that the rest went below the main beam (sorry no "below" picture).
So, to recap the history of the beam quality,
Other things.
Just to make sure, we turned down the 9MHz and 45MHz RF power to 3dBm and disconnected the 118MHz and 24MHz cables and nothing changed.
We know that the crystal wedge is supposed to be horizontal and we know that the wedge orientation is correct. When we first installed the EOM in chamber, the EOM transmission was deflected horizontally in +Y direction.
Curoius thing about the EOM dimensions
Crystal length L=40mm, thickness T=4mm=L/10, wedge angle w=2.85 deg, and tan(2w) = 0.09981 ~ 1/10.
Though this is probably not related to the ghost beams in PIT direction, when the beam is perfectly aligned with the EOM (i.e. the light traveling the center of the crystal), the internal AR reflection of at the output face of the crystal hits the side of the crystal and the specular reflection will hit the input surface of the crystal and almost exactly comes back on top of the main beam with only 0.0272mm offset. See the 1st cartoon.
Note that the side surfaces are not polished (though the AOI is 84.3 deg so most of the power is reflected back into the crystal due to total internal reflection).
If you displace the beam in horizontal direction, the AR path is displaced in the opposite direction by about the same amount (i.e. if the main beam moves by 0.5mm toward the short face of the crystal, the AR-side-AR beam moves by about 0.5mm toward the long face). If you continue tracing the AR-side-AR beam, it turns out that the AR-side-AR-AR-side-AR beam will come back exactly on the main beam. See the 2nd cartoon (which is actually to scale, the main beam is off by 0.5mm and the 1st ghost is off by 0.5272 in the opposite direction, and the 2nd ghost is on top of the main beam).
Interesting design choice.
Following on from this log (alog #89046), we have already checked that the MC mirrors are restored to their previous positions from before the power cut in December. We know the seismic state now should be as close as we can manage to this time 2025-12-03 11:28:44 UTC when the mode cleaner was locked at 2W input power and before any HEPIs were locked for the vent.:
Summary of seismic state now:
HAM ISI locked. HAM 2, 3, 4, 5, 6 ISI isolated.
HAM1,2 and BSC2 HEPI locked. HAM 3,4,5,6 HEPI isolated - ie. as per nominal operation.
I trended the IM osems and the MC2 and IM4 QPDs in HAM2 while the mode cleaner was locked for mode scans yesterday (reference time is 2026-02-05 18:44:36 UTC).
It seem as though the MC2 and IM4 trans powers we have are lower now, but also the IM mirrors have changed.
As Sheila did yesterday I will try and summarise these changes in a table and compare whether this makes sense with the greater HAM1 loss we have because of the JAC (~40 %).
Just posting an updated version of Sheila's table to compare this new time (3rd December).
| 2025-12-03 | 2026-02-05 | ||
| seismic state | nominal | HAM1 ISI, HEPI locked, HAM2 HEPI locked. | MC mirrors should be at or close to nominal assuming table alignment didn't change much. |
| power into HAM1 H1:IMC-PWR_IN_OUT16 | 2W | 1W | |
| MC2 trans | 310 counts | 90 counts | consistent with 60% HAM1 throughput measured with a power meter (extra loss due to temporary mirror for JM3) |
|
IM4 trans nsum |
1.8W | 0.1W | should be 0.54 W so not consistent with JAC loss. |
| IM1 position P, Y | 3130, -679 | changed by -11, -0.44 | Off in pitch |
| IM2 P, Y | 915, -226 | -50, -3 | Off in both, mainly pitch. |
| IM3 P, Y | 44, -1634 | +4, -13 | Off in both, mainly yaw |
| IM4 P, Y | -2815, -22 | +64, -8 | Off in both, mainly pitch. |
So after this Sheila and I are going to try turning on the IMC ASC and working out if we can tune the alignment onto IM4-TRANS using the IMs.
Since the IMC is nicely locked, and we're only seeing 0.2 W on IM4 trans at a time when we expect ~1W, I moved the IM1, IM2, and IM3 to these positions. Still not very much on IM4_Trans.
While Keita will report the detail about the heater wiring trouble shooting, here is the quick report for the JAC heater functionality.
The attatched plot shows the two thermistor signals (top/bottom) and the heater input (middle). The temperature is reasonably changing with the heater driver input.
JAC in-vac tri-cable was swapped. (Masayuki, Jennie, Jason, Keita)
The cable that failed (D2500336-V2, S2501241) was pulled out of the chamber. (In the process, one of the body mode damper crossbar had to be temporarily removed to release the cable. ) The cable was wrapped and put in a bag without further testing (yet).
Jennie and I tested the new tri-cable (D2500336-V2, S2501242) in the optics lab and it was good (i.e. every pin was connected to the pin it is supposed to be connected, no cross-wiring, no short circuits).
The new cable was installed in chamber.
I checked the in-chamber connection from the in-air side of the D4F10 feedthrough. Pin13 wasn't connected to anything, ditto for the chamber ground, thermistors showed about 11kOhm each, heaters were about 50 Ohm each, PD anode and cathode were good, no cross-wiring and no unintended short circuits.
After connecting the in-air cable to the feedthrough, PZT and Trans PD worked right away.
After connecting the heater cable to the driver chassis, we confirmed (using the breakout board and a DVM) that the voltage across the heater elements was ~1.4V when H1:JAC-HEATER_DRV_VSET~3.6[V?], H1:JAC-HEATER_DRV_VMON~1.6[V?] and H1:JAC-HEATER_DRV_IMON~-0.098[A?]. Maybe 1.4V is close enough to 1.6V (0.2V might be the voltage drop of the in-air cable?) but I don't understand the logic about VSET being 3.6V nor why IMON is negative (in my mind it seems logical if IMON~1.6V/25Ohm = 0.064A).
But it's good to know that it's doing something.