H1 CAL (AOS, CAL)
sudarshan.karki@LIGO.ORG - posted 19:32, Wednesday 29 July 2015 (20054)
PCALX Clipping Issue (may have been solved)

Sudarshan, Travis

We used today's maintenance  period to check the PCALX clipping issue mentioned in  alog #19899. We moved the clipped beam  using the last mirror ( the one that directs the light to the test mass) in the transmitter module  and were able to get the reflected beam on the receiver side at a fairly reasonable level.  There was  a very narrow window in our alignment in  both pitch and yaw to get the beam out on the receiver side and we are guessing we are very close to the edge of one of the optics inside the chamber.

The laser power as measured by ophir power-meter are as  follows:

  Inner Beam Outer Beam
Transmitter Side 0.738 W 0.734 W
Receiver Side 0.718 W 0.712 W

We will check the beam balance between the two beams and optical efficiency during our next calibration and touch up the alignment if needed.