Reports until 15:52, Friday 31 July 2015
H1 AOS
david.barker@LIGO.ORG - posted 15:52, Friday 31 July 2015 - last comment - 15:56, Friday 31 July 2015(20107)
attenuated IRIG-B signal at end stations, rebooted h1pemmy

Robert, Vinny, Stefan C, Dave:

The recently installed IRIGB signals were clipping the ADC due to the 10X gain in the PEM AA chassis. We installed a 10X attenuator in the line at the IRIG-B chassis in both end stations, the IRIGB signal range is now 0 - 5000 counts.

We also power cycled h1pemmy to see if this would remove the 64Hz noise being seen in the SEIS channels there. The procedure was: stop all models, power down h1pemmy, power down IO Chassis, power up IO Chassis, power up h1pemmy. We got lucky on the auto-code start and the timing came back OK with no further IOP restarts required.

Unfortunately this power cycle does not seem to have fixed the noise.

Comments related to this report
david.barker@LIGO.ORG - 15:56, Friday 31 July 2015 (20108)

The attenuated IRIGB signals are shown, first 1/16th of the second at 16kHz.

Images attached to this comment