Displaying report 1-1 of 1.
Reports until 20:39, Sunday 02 August 2015
H1 ISC (ISC)
stefan.ballmer@LIGO.ORG - posted 20:39, Sunday 02 August 2015 (20151)
Broadband frequency noise injection during thermal tuning (sneak preview)
Attached is a plot of the frequency noise in OMC_PDC_A at six different times during the thermal tuning. There is a significant drop in frequency noise coupling. However the feature that changes most drastically is the carrier 10-mode resonance at 5.1kHz. Thus some of this might be alignment induced effects.

Plot 2 has three more traces with even more heating.

Images attached to this report
Displaying report 1-1 of 1.