H1 CDS (ISC)
jameson.rollins@LIGO.ORG - posted 09:58, Monday 03 August 2015 - last comment - 16:48, Monday 03 August 2015(20157)
h1lsc test points cleared to clear up errant excitation in

There was a persistent excitation in:

H1:ALS-C_REFL_DC_BIAS_EXC

on h1lsc.  This was presumably from an awg process left running by Evan on operator1 workstation.

I killed ALL testpoints on h1lsc (dcuid 10):

jameson.rollins@operator1:~ 0$ diag -l -z
supported capabilities: testing  testpoints  awg 
diag> tp clear 10 *
test point cleared
diag>

Comments related to this report
james.batch@LIGO.ORG - 16:48, Monday 03 August 2015 (20175)
Actually, you should clear the awg, then clear the test point. As it is now, there is still an awg slot with a sine excitation, but no test point.