Added a few tests today and removed ITMX/Y L2 OLDAMP Off notification.
Tests added are:
-
ESD railed - if the absolute value of the average after 5 seconds stays within +/- 600 of 16200, then it will notify.
-
TCS power - if the power is < 50W, notify.
-
TCS Flow and IR alarms - if in alarm, notify.
-
Beckhoff down - if the elasped time is not greater than the last time the test ran, then it will notify (trending showed that this channel will stop counting when Beckhoff is crashed)
As always, please be aware of these notifications and always investigate when one pops up.