Displaying report 1-1 of 1.
Reports until 20:14, Tuesday 04 August 2015
H1 TCS (SYS)
evan.hall@LIGO.ORG - posted 20:14, Tuesday 04 August 2015 - last comment - 06:52, Wednesday 05 August 2015(20233)
Rotation stage roulette

Patrick, Jeff, Evan

We spent a few minutes cooking the IX compensation plate while trying to make the TCS rotation stage behave.

Images attached to this report
Comments related to this report
patrick.thomas@LIGO.ORG - 20:26, Tuesday 04 August 2015 (20236)
This was after the Beckhoff chassis was power cycled.
matthew.heintze@LIGO.ORG - 06:52, Wednesday 05 August 2015 (20248)

If after the beckhoff chassis is cycled and if you do not first "refind home" then you will find these issues with the rotation stage not knowing what angle to go with as it loses its mind where its at. I cant say I have noticed the waveplate acting up after this step is taken (but maybe it has without me noticing)

 

Also note that search for home does not take the waveplate to minimum angle necessarily. You should be pressing "go to minimum power". So after a Beckhoff restart, usually here at LLO we search for home first, so that the rotation stage finds its 0 point again, then go to minimum power and then start operating it from there. A known issue all along is that as the waveplate rotates to home it could go through a brief period where it allows maximum power into the IMC or onto the ITM CPs

Displaying report 1-1 of 1.