Summary:
I looked at all 10 events on Gabrielle's alog that are without "Those marked with ** look somwhat different in the spectral shape" caveat, i.e. I only picked the events that were similar according to Gabrielle.
We don't know what these are, we know that this is NOT a sensing side software/hardware glitch.
All of these look very similar (see attached). DCPD current suddenly shoots up, there's a msec-ish rise, and then an equally sudden and fast fall of msec order, then the servo slowly follows up.
Details:
The spikes for larger glitches are on the order of 10^-15m to 10^-14m, or 0.1%-ish RIN (0.02mA) (but there are smaller ones).
DC current increases, which means that X becomes shorter and/or Y becomes longer (if the IFO calibration is correct).
What we were able to exclude so far is:
Things that are not yet completely killed are:
Is there anything that helps us understand if this is happening inside or outside of the arms?
Plots:
These are four biggest glitches from Gabrielle's list but I looked at all 10 that are without "look somewhat different" remark.
On top is the calibrated DARM displacement, properly dewhitened, and rewhitened again to make the glitch visible while keeping the phase intact above 600Hz or so, and mildly band-stopped (-20dB) second violin resonances. The "calibration" of this is 10^-10 meters, and the sign is correct for msec signal (i.e. positive means X-Y going positive).
Middle is the OMC DCPD SUM. DARM UGF is about 40Hz (1/e time = 1/2/pi/40=4msec), so the glitch shape is almost undisturbed by the DARM servo.
Bottom (of the bottom) is the shape of the glitch after subtracting the background motion in OMC DCPD SUM.
Some details on the Virgo glue glitches can be found at the following elog entries:
WI glitch rate: shows the time domain shape of some of the glitches, that were quite fast kicks
Glitches: Another example time series
Glitch: where the excitation of one of the test mass modes is evident
Some Virgo jargon translation
Pr_B1_ACp = DARM error signal, equivalent to AS_RF_Q
Hrec = calibrated (high passed at ~10 Hz) strain, equivalent to CAL_DELTAL_EXTERNAL
WI = West input mirror ITMY