kyle.ryan@LIGO.ORG - posted 17:21, Friday 21 August 2015 - last comment - 18:17, Friday 28 August 2015(20765)
Beam Tube ion pump coupled to port hardware at Y2-8
Kyle, Gerardo 

Today we coupled the ion pump to Y2-8 -> The operation was slow and meticulous but we feel confident that no new net forces are realized by the beam tube nozzle -> We also pumped out and leak tested -> Still to do is the pulling of the HV cable and bake out -> In the meantime we will likely move on to X2-8 and repeat the exercise.  

(Note that Mike Z., Ken Mason and others provided the components used)
Non-image files attached to this report
Comments related to this report
kyle.ryan@LIGO.ORG - 18:17, Friday 28 August 2015 (21000)
Upon reconsideration the BT valve is experiencing a minimum of 55ft*lbs as a result of having mounted the 8" gate valve to the reducing Tee before having coupled the ion pump to the 8" valve