Displaying report 1-1 of 1.
Reports until 13:04, Wednesday 25 January 2012
X1 SUS
jeffrey.garcia@LIGO.ORG - posted 13:04, Wednesday 25 January 2012 (2083)
X1 SUS QUAD 03 BLD 03 M0 & R0 TFs
X1 QUAD 03 BUILD 03

The first set of matlab TFs for M0 & R0 masses on the latest QUAD build (QUAD 3 BUILD 3) with full cable lacing to the lower stages.  The M0 measurement agrees very well with the model for all DoFs, with the Pitch DC adjustment a little lower than the model (as expected from every previous measurement).  For the R0 measurement, the L and T second modes are slightly lower in frequency from the model and with the same lowered Pitch DC level discrepancy, although it is a bit more than the M0 difference.  Comparisons with previous measurements to follow.
Non-image files attached to this report
Displaying report 1-1 of 1.