Added a gain change for ASAIR_B_RF18, ASAIR_B_RF90, ASAIR_B_LF and ASAIR_A_LF when the AS port beam diverted closes (it has 10% residual through beam).
# Set the ASAIR PD gains for BDIV open
ezca['LSC-ASAIR_B_RF18_I_GAIN'] = 18
ezca['LSC-ASAIR_B_RF18_Q_GAIN'] = 18
ezca['LSC-ASAIR_B_RF90_I_GAIN'] = 18
ezca['LSC-ASAIR_B_RF90_Q_GAIN'] = 18
ezca['LSC-ASAIR_A_LF_GAIN'] = 11
ezca['LSC-ASAIR_B_LF_GAIN'] = 11
Curiously, the DC gains scale differently than the RF channels...