Attached are ASD and Coherence betwix the three corner station STS2s taken at 3am local Wednesday Thursday and Friday. These three days look very similar. I installed STS2-A aka HAM2 on Tuesday. Recap:
STS2-C (HAM5) has been in place for long time. STS2-B (ITMY) is the unit returned from Stanford replacing PEM STS2 which we were using to replace our original STS2-B which is still at Quanterra for repairs. STS2-A returned from Quanterra after a couple months in their vault where they say there was no problem. The A unit spent time in the BeirGarten near the ITMy unit undergoing cable and chassis swaps, see 18354 for recap but nothing was consistent.
To me, STS2-A still looks like this unit has a problem. When I look at plots from back in May, 18354, it looks even worse now. There was better coherence in Z and X with the other units and Y was equally poor.
RobertS--do you have an opinion?
Ignore the Title stating Huddle and cables switched. Units are in home position on nominal cables.