Reports until 22:46, Monday 31 August 2015
H1 SUS (CDS, SUS)
cheryl.vorvick@LIGO.ORG - posted 22:46, Monday 31 August 2015 - last comment - 23:52, Monday 31 August 2015(21077)
EVE OPS shift: the story so far, and current plans

shift started with PRM satellite box swaps (8/31 23:00 - 9/1 00:10UTC, one hour)

 

relocking issues: 9/1 00:10UTC to 5:14UTC

- alignment that I corrected without a full realignment (9/1 ended 1:00UTC)

- delays by BS ISI being tripped - happened twice (ended 2:14UTC)

- delayed by SR3 glitching, preventing IFO from going to CARM_5PM (2:14UTC to 5:14UTC)

 

Having called Vern, Richard, Dave: 5:14UTC

 

Dave restarted h1nds1, so it's back

- option later tonight is to power cycle the computer and then log in and restart the process

Richard

- options are to power cycle AI chassis and or coil drivers for SR3

- and then change the coil driver if power cycle doesn't work

Vern

- ok to call off PWL shift if no SR3 fix and therefore no locking 

- have emailed Jim who's on OWLs, to let him know

 

Currently:

Sheila and Jenne out to LVEA to power cycle AI chassis and coil driver on SR3 (5:46UTC)

Comments related to this report
cheryl.vorvick@LIGO.ORG - 23:31, Monday 31 August 2015 (21078)

rebooted AI and coil driver - no fix

- no change

 

powering off AI for a few minutes 6:08:11UTC

- voltmon power spectra shows T2 noise is now gone

 

AI chassis swapped out 6:24UTC

- SR3 looks better, going to CARM_10PM to look at it there

cheryl.vorvick@LIGO.ORG - 23:52, Monday 31 August 2015 (21079)

Anamaria's comment about voltmon:

when moving SR3 at DC using the alignment sliders we see 1urad of pitch on M0 shows up as 0.5 urad pitch of oplev and 2 cts of voltmon.