Earlier this month, we had made a complete assessment of all aLIGO SUS signs and flips across suspension types (see T1200015) and currently functional builds of those types (see SignStatus_120111.pdf).
In doing so we identified:
- Some suspensions had a few of their magnets flipped
- The BSFMs had some mistakes in their Basis Transformation Matrices
- Some suspensions had there OSEMINF sat-amp compensation filters defined with more poles than zeros
For the most part, we have now rectified these problems. I attach a new assessment. The remaining action items are
- Fix H2 ITMY M0's F2 and F3 magnet sign; to fix when we swap out for real ITM Optic
- Figure out what's wrong with X2 PR3's SD OSEM; currently under investigation.