We were able to be back to full lock at around 17:00 UTC. There were a couple of notable points as listed below:
-
End Y fiber polarization had too high wrong polarization
-
TJ, Corey and I adjusted the polarization in the MSR. I accidentally set the X arm polarization to default, so we did it on X arm too. Now they are good.
-
AS90 offset changed again.
-
I adjusted the dark offset. I am not sure if this is related to my activity around the rack by HAM6 (see alog 21101)
-
ISC_DRMI guardian had a problem
-
The screen turned in dark red with an error message like "epics data was not read" or something similar.
-
A funny thing is that it kept running main of the ASC_ENGAGE more than 10 times, not just twice or so.
-
Manually re-requesting ASC_ENGAGE fixed the problem.
-
ISS 2nd loop engage failed once
-
It did not unlock the interferomter, but gave a big glitch in the arm power and elese where.
-
We manually requested PREPARE_ISS and let it sit there for a while ~ a minute. Then requesting ISS_ON made it without a glitch.