Displaying report 1-1 of 1.
Reports until 09:50, Monday 14 September 2015
H1 PSL
jim.warner@LIGO.ORG - posted 09:50, Monday 14 September 2015 (21503)
PSL Status
		Laser Status: 
SysStat is good
Front End power is 33W (should be around 30 W)
FRONTEND WATCH is GREEN
HPO WATCH is RED

PMC:
It has been locked 12 days, 23 hr 3 minutes (should be days/weeks)
Reflected power is 2.1 Watts  and PowerSum = 26.5 Watts.
(Reflected Power should be <= 10% of PowerSum)

FSS:
It has been locked for 2 h and 24 min 
TPD[V] = 1.48V (min 0.9V)

ISS:
The diffracted power is around 8% (should be 5-9%)
Last saturation event was 2 h and 51 minutes ago 

NOTES: 
Displaying report 1-1 of 1.