H1 DetChar (DetChar, ISC)
andrew.lundgren@LIGO.ORG - posted 01:24, Tuesday 15 September 2015 - last comment - 04:12, Tuesday 15 September 2015(21526)
Short glitches about every half-second, only in DARM
This 2 Hz comb is actually glitches every half a second. It seems like the period is not exactly half a second. They appear in DARM but not MICH, SRCL, or PRCL, so it's probably not a bad PR or SR mirror driver.

Attached are a few Omega scans. The first is several seconds long and shows that these are very short and span about 40 to 100 Hz. The next three Omega scans are 4 seconds apart, showing that these don't sit right on the integer second and have a period that is not quite 0.5 seconds.
Images attached to this report
Comments related to this report
thomas.abbott@LIGO.ORG - 03:15, Tuesday 15 September 2015 (21527)
The comb shows up in the ETMY top stage F1, F2, and SD but none of the others. See attached plots.
Images attached to this comment
thomas.abbott@LIGO.ORG - 04:12, Tuesday 15 September 2015 (21528)
Nutsinee, Thomas A.,

Ran omega scan for 1126336348 (Sep 15 2015 07:12:11 UTC), and found that the once-per-0.5s glitches repeating glitches are noticeable in:

H1:OMC-DCPD_SUM_OUT_DQ

H1:ISI-ETMY_ST2_BLND_RZ_GS13_CUR_IN1_DQ

H1:ISI-ETMY_ST2_BLND_Z_GS13_CUR_IN1_DQ

H1:PEM-EY_MAG_EBAY_SUSRACK_QUAD_SUM_DQ

H1:PEM-EY_MAG_EBAY_SUSRACK_X_DQ

H1:PEM-EY_MAG_EBAY_SUSRACK_Y_DQ

H1:PEM-EY_MAG_EBAY_SUSRACK_Z_DQ

Full omega scan: https://ldas-jobs.ligo-wa.caltech.edu/~tabbott/H12HzComb/
Images attached to this comment