The full report can be found on the detchar wiki, but here are the main highlights:
-
Nice long locks. Duty cycle 72%. Majority (5/9) of lock losses seem to be due to earthquakes
-
Main downtime during these 4 days has been due to Chile earthquake which took place just before the start of the shift (16th at ~23:30 UTC) and subsequent aftershocks.
-
The 48/49 Hz glitches seem to be becoming more prominent; hveto picks out PEM-EX_ACC_BSC9_ETMX_Z_DQ, HPI-ETMX_BLND_L4C_RX_IN1_DQ, HPI-ETMX_BLND_L4C_Y_IN1_DQ channels as good witnesses
-
The 60 Hz glitches are continuing; hveto picks SUS-ETMY_M0_DAMP_Y_IN1_DQ and SUS-ETMY_L2_WIT_L_DQ channels as good witnesses
-
300-400Hz glitchy region (periscope?) is much worse at the beginning/end of a lock stretch. Also noticing this region more around the times of earthquakes/seismic activity (which is not so surprising). ASC channels are good witnesses of this region.
-