Displaying report 1-1 of 1.
Reports until 02:12, Thursday 24 September 2015
H1 General
corey.gray@LIGO.ORG - posted 02:12, Thursday 24 September 2015 - last comment - 02:16, Thursday 24 September 2015(21879)
H1 46hr Lock Ended & Back Up

H1 had a ~46Hr lock that ended toward the end of Jim's shift.  He noted DRMI was looking pretty ugly.  I waited DRMI a little, but eventually gave Sheila's new PRMI procedure a try, but POP18 & POP90 remained at a flat zero and PRMI never locked.  So went through the Initial Alignment procedure (I went slow & took notes since this is something we do rarely.).  Alignment was and getting to NOMINAL_LOW_NOISE were straightforward.

Reason for Lockloss?

I saw nothing obvious from seismic signals when I walked in, everything quiet!  Jim mentioned ASAir/OMC video spots took an odd move prior to lockloss.  H1 was trending down over 6hrs before the lockloss, and Jim mentioned this could be the RF45 issue which has been coming up.

Initial Alignment Notes:

Comments related to this report
corey.gray@LIGO.ORG - 02:16, Thursday 24 September 2015 (21880)

Accidental Slip Out Of Observation Mode:  9:04:15 - 9:04:33

While making the entry above and updating the Initial Alignment wiki, I wanted to look at a pull-down of an ISC Guardian node.  Clicking on the ALS_YARM node pull-down dropped H1 out of Observation & into Commissioning.  (after cursing, I quickly checked things and took H1 back to Observation)

(I didn't bother touching the Observatory Mode button during this short slip.)

Displaying report 1-1 of 1.