Displaying report 1-1 of 1.
Reports until 16:20, Thursday 24 September 2015
H1 INJ (DetChar, INJ)
christopher.biwer@LIGO.ORG - posted 16:20, Thursday 24 September 2015 - last comment - 17:45, Thursday 24 September 2015(21911)
Beginning CBC hardware injection test
We have tinj running again at LHO. We are going to test scheduling injections again.
Comments related to this report
christopher.biwer@LIGO.ORG - 16:24, Thursday 24 September 2015 (21913)DetChar, INJ
1127172657 1 0.5 coherenttest1_1126257410
christopher.biwer@LIGO.ORG - 16:27, Thursday 24 September 2015 (21914)
tinj has died again near 23:27:24 UTC.
christopher.biwer@LIGO.ORG - 16:37, Thursday 24 September 2015 (21915)
tinj reported missing injection file for the previous injection. Filename needed a _H1. and schedule need a _ at the end of the filename.

A new test with changed filename.

1127173421 1 0.5 coherenttest1_1126257410_
christopher.biwer@LIGO.ORG - 16:57, Thursday 24 September 2015 (21917)
The injection schedule at 1127173421 was successful with tinj. Attached is the omegascan.
Images attached to this comment
christopher.biwer@LIGO.ORG - 17:15, Thursday 24 September 2015 (21919)
1127175848 1 1.0 coherenttest1_1126257410_
christopher.biwer@LIGO.ORG - 17:29, Thursday 24 September 2015 (21920)
The injection at 1127175854.0 was successful.
Images attached to this comment
christopher.biwer@LIGO.ORG - 17:45, Thursday 24 September 2015 (21922)DetChar, INJ
We have scheduled three more test injections:

1127179817 1 1.0 coherenttest1_1126257410_
1127181617 1 1.0 coherenttest1_1126257410_
1127183417 1 1.0 coherenttest1_1126257410_

These are the last of the night. I will report the events of tonight in another aLog entry.